Dissolved Oxygen Calibration
Single Point Calibration
Press the “Single-pt” button.
Remove the cap and discard it. The cap is only used for protecting the probe during shipping.
Expose the probe to air for 5-30 seconds.
Press the “Calibrate” button. The screen should show “Successful Calibration;" otherwise, retry the calibration.
Dual Point Calibration (Optional)
Press the “Dual-pt” button.
Remove the cap and discard it. The cap is only used for protecting the probe during shipping.
Press the “Calibrate” button. The screen should show “Successful Calibration;" otherwise, repeat steps 1-3.
Place the probe in the Dissolved Oxygen solution, and stir to remove trapped air.
Let the probe sit in the solution for 10-90 seconds.
Press the “Calibrate” button. The screen should return to the main screen for Dissolved Oxygen; otherwise, repeat steps 4-6.